sunbet(中国)

    CN/ EN

    PSEM Series

    PSEM Series

    PSEM Series Room Temperature Electromagnet Probe Stage is a probe stage specially developed for room temperature variable magnetic field electrical measurement environment, providing ±0.6T vertical magnetic field or ±0.3T horizontal magnetic field, capable of 2 inch, 4 inch wafer repeatability, standard electrical experiments, external different test equipment can be completed on the device's electrical characteristics, parametric measurements, DC measurements, RF measurements, and Hall measurements.

    PSEM Series Overview

      The Room Temperature Electromagnet Probe Stage not only performs the tests that can be done with a permanent magnet probe stage, but also provides a variable magnetic field for sample testing with a stronger magnetic field and a larger uniform magnetic field zone. This Probe Stage is designed for experiments that require continuous magnetic field variation during the measurement process and non-destructive testing of chips, wafers and devices at room temperature.


      PSEM Series Room Temperature Electromagnet Probe Stage is a probe stage specially developed for room temperature variable magnetic field electrical measurement environment, providing ±0.6T vertical magnetic field or ±0.3T horizontal magnetic field, capable of 2 inch, 4 inch wafer repeatability, standard electrical experiments, external different test equipment can be completed on the device's electrical characteristics, parametric measurements, DC measurements, RF measurements, and Hall measurements.


    Features:
    - The sample holder can hold 4-inch wafer samples. By moving the slide under the sample holder, it can achieve ±50mm travel in X-Y axis, and the sample holder itself can be finely adjusted in three dimensions, which makes the sample testing and sample changing more convenient.
    - The base of the probe arm is fixed with a magnet, allowing the probe to be adjusted in three dimensions, X-Y-Z, and with the sample holder's shift adjustment, the probe can be quickly inserted into any position of the 4-inch sample (6 probe arms can be mounted, and 4 horizontal magnetic field electromagnets can be mounted in the probe table).
    - Probe arm with triaxial cable and triaxial connector, small leakage, leakage current within 100fA, probe arm built-in cable to avoid messy wiring.
    - The probe is fixed with needle sleeve, only the tip of the needle is exposed, reducing the leakage of electricity.
    - Electromagnet bracket and probe stage adopt separate isolation bracket to avoid vibration transmission to the sample when the magnet is moving.
    - Sample fixing method adopts porous partition adsorption, there are three individually controlled gas adsorption channels in the outer ring, inner ring and the middle, which can adsorb and fix 1mm*1mm samples.
    - CCD magnification of 180 times, working distance of 100mm.

    PSEM Series Technical Parameter

    Model classification:
    Model PSEM-2H(Horizontal Magnetic Field)
    PSEM-2V(Vertical Magnetic Field)
    PSEM-4V(Vertical Magnetic Field)
    Sample seat size2 inches2 inches4 inches

    Parameters and indicators:

    Sample holder
    Chuck size
    2/4 inches, customizable
    Material
    Oxygen-free copper gold plating
    Sample fixation method
    The porous zonal adsorption has three separately controlled gas adsorption channels: the outer ring, the inner ring and the middle
    滑台
    X-Y movement stroke
    2/4 inch :±50mm (Horizontal magnetic field probe station without slide table)
    Movement accuracy
    10μm
    探針臂
    X-Y-Z movement stroke
    X-13mm,Y-13mm,Z-13mm
    Degree of rotation
    360-degree free rotation
    Movement accuracy
    10μm
    Number of probe arms
    The vertical magnetic field electromagnet probe station can be installed in six units. Four horizontal magnetic field probe stations can be installed
    Electrical cablesThree-coaxial cable
    Leakage current
    100fA
    Probe type
    Dc probe
    Optical system
    Microscope magnification
    10 to 180 times
    Working distance of the microscope
    90-100mm
    CCD camera resolution
    38 million pixels
    Electromagnet assembly (vertical magnetic field)
    Magnet model
    EM-60V
    Magnetic field magnitude
    ±0.6T
    Magnet spacing
    30mm
    Pole post
    The diameter is 60mm, and it comes standard with a pair of 50mm pure iron terminals
    Circulating water cooling
    Electromagnet assembly (horizontal magnetic field)
    Magnet model
    EM-30H
    Magnetic field magnitude
    ±0.3T
    Magnet spacing
    60mm
    Pole post
    It comes standard with a pair of 60mm pure iron terminals
    Circulating water cooling
    Optional
    Optical platform/DC probe/microwave probe/sample holder/shielding box/optical fiber


    PSEM Series Basic Configuration

    Semiconductor
    IC
    Wafer

    recommend